Mini Data Logger For Voltage (50V) LR5043
Hioki compact 1 or 2-channel data loggers feature a 60,000 data set/channel memory to measure temperature, humidity, voltage or instrumentation signals. The LR504X series logs DC voltage over 1 channel in as fast as 1 second intervals, available in three different ranges.
Key Features
- Easily mount the light-weight , pocket-sized loggers in tight spaces
- Easy-to-see dual display
- Transfer data to PC even during recording
- Replace batteries while recording (30 second limit)
- 3 times the memory capacity than predecessor (Record 60,000 data per channel)
- Record without missing fluctuations in STAT mode
- Measurement data is preserved even after the battery dies - Worry-free backup preserves recorded data even if a new measurement is started by mistake.
Product Warranty Period: 3 years
Included Accessories:
LR6 (AA) Alkaline battery (built-in internal) ×1, Connection cable LR9802 ×1, Instruction manual ×1, Operation guide ×1, Kickstand ×1
LR5041 | LR5042 | LR5043 | |
Measurement items | DC voltage 1ch | DC voltage 1ch | DC voltage 1ch |
Measurement range | -50.00 to 50.00 mV | -5.000 to 5.000 V | -50.00 to 50.00 V |
Accuracy | ±0.5 %rdg. ±5 dgt. | ||
Storage capacity | Instantaneous value mode: 60,000 data, Statistical value mode: 15,000 data | ||
Recording interval | 1 to 30 sec., 1 to 60 min., 15 selections | ||
Recording modes | Instantaneous recording: at every recording interval Statistical value recording: Measure at one second intervals, and record the instantaneous, maximum, minimum, and average values within every recording interval | ||
Recording methods | One-time recording: Stop recording when the memory capacity is full Endless recording: Continue recording even when the memory capacity is full (old data is overwritten) Start: Logger button operation or scheduled time Stop: Logger button operation or scheduled time, or auto-stop when the memory capacity is full (at one-time recording) | ||
Other functions | Pre-heat function (requires external power supply during use of function), Always backs up last recorded data; backs up recorded data and setting conditions when battery power is low; guarantees approx. 30 sec. of recording operation and clock while battery is replaced | ||
Waterproof and dust-proof | IP54 (EN60529) (with connection cable connected, but not including cable tip) | ||
Interfaces | Infrared optical communications with LR5091, LR5092-20 | ||
Power supply | LR6 (AA) Alkaline battery ×1, Battery life: Approx. 2 years (Instantaneous recording, with 1-minute interval and auto power saving, at 20 °C), Approx. 2 months (Instantaneous recording, with 1-second interval at 20 °C) | ||
Dimensions and mass | 79 mm (3.11 in) W × 57 mm (2.246 in) H × 28 mm (1.10 in) D, 105 g (3.7 oz) | ||
Included Accessories | LR6 (AA) Alkaline battery (built-in internal) ×1, Connection cable LR9802 ×1, Instruction manual ×1, Operation guide ×1, Kickstand ×1 |
ACCESSORIES
CONNECTION CABLE LR9802
- Tips 4 wires, 1 m (3.28 ft) lengthCOMMUNICATION
ADAPTER LR5091
- Transfer data from logger to PC
- Transfer setting/clock data from PC to logger
- Free bundled software
WALL-MOUNTED HOLDER LR9901
- Not compatible with Model LR5051
MAGNETIC STRAP Z5004
- Strap with magnet
DOWNLOADS
Catalog
- DATA LOGGER LR5000 Series
Manual
- VOLTAGE LOGGER (50mV) LR5041,(5V) LR5042,(50V) LR5043
Declaration of Conformity